作 者:任自文[1] 丁燕生[1] 陈健[1] 吴林[1] 张筠华
机构地区:[1]北京医科大学第一医院心内科
出 处:《中华心血管病杂志》1997年第6期442-445,共4页Chinese Journal of Cardiology
摘 要:国内首次合并应用埋藏式心脏复律除颤器(ICD)与单极起搏器1例。术中仔细观察起搏器对ICD的影响,未发现ICD对起搏心律的双感知,亦未发现对心室颤动(室颤)的感知不足,ICD对3次诱发的室颤均迅速感知并一次除颤成功。术后189天储存资料显示患者共自发21次室性心动过速,均被一次抗心动过速起搏有效终止,未发生误放电,亦未漏诊快速心律失常,初步观察效果满意。This is the first case combining use of implantable cardioverterdefibrillator (ICD) with unipolar pacemaker in our country. The effect of pacemaker on ICD was carefully evaluated during implantation. There were no double sensing of ICD to pacing rhythm and no undersensing to ventricular fibrillation (VF). ICD rapidly detected the induced VFs in three tests and successfully defibrillated the Vf by one shock in every test. The stored data of 189 days postimplantation showed that total twenty one episodes of spontaneous ventricular tachycardias were detected and every episode was successfully terminated by only one burst of antitachycardiac pacing. There was no faulty delivery of therapy and no loss of detecting tachyarrhythmias. The result of combined ICD with unipolar pacemaker was satisfying in the preliminary observation.
关 键 词:人工心脏起搏 心动过速 心室颤动 ICD 起搏器
分 类 号:R541.71[医药卫生—心血管疾病]