作 者:吴玲玲[1] 秦芳[1] 黄爱军[1] 顾光煜[1]
机构地区:[1]江苏省南京大学医学院附属鼓楼医院输血科,210008
出 处:《临床输血与检验》2008年第4期300-302,共3页Journal of Clinical Transfusion and Laboratory Medicine
摘 要:目的观察Sysmex XS800i血液分析仪提示白细胞异常的可信程度。方法对血液分析仪提示白细胞异常信号的标本同时进行显微镜镜检,并对比分析。结果Sysmex XS800i血液分析仪对正常的成熟白细胞分类已达到较理想的程度,对异型淋巴细胞和原始、幼稚细胞的提示,与显微镜镜检结果符合程度低于成熟细胞。结论血液分析仪不能完全替代显微镜镜检,一旦仪器有异常细胞警告提示或临床高度怀疑细胞形态异常时,必须作涂片显微镜镜检。Objective To study the confidence level of indicating abnormal leukocyte by Sysmex-XS800i blood analyzer. Methods When specimens were indicated by Sysmex-XS800i blood analyzer as abnormality,they were contrastively inspected by microscope simuhaneously. Results Using Sysmex-XS800i blood analyzer on classifying normal maturity leukocyte was considered ideal, but was unfit for abnormal lymphocyte such as primitive and immature cells when compared with microscope test. Conclusion Blood analyzer couldn't replace microscope test completely. When abnormal cells were indicated by blood analyzer or suspected by clinician,these specimens must be smeared for microscope test.
关 键 词:SYSMEX XS800i血液分析仪 白细胞异常 显微镜镜检