火焰原子吸收光谱法同时测定尿液中钾和钠  被引量:5

Simultaneous Determination of K and Na in Urine with FAAS

在线阅读下载全文

作  者:易海艳[1,2] 胡建安[1] 

机构地区:[1]中南大学公共卫生学院,湖南长沙410007 [2]湖南省劳动卫生职业病防治所,湖南长沙410007

出  处:《精细化工中间体》2008年第5期68-70,共3页Fine Chemical Intermediates

摘  要:采用原子吸收光谱仪同时测定了尿液中钾和钠的含量,并确立了测定钾和钠的优惠条件。加入氯化铯可消除电离干扰,当稀释比例达到1:1000时,可以消除尿基质的干扰。钾和钠的检测限分别为0.0049mg/L和0.0040mg/L,测定范围分别为0.2-5.0mg/L和0.2-3.0mg/L,方法回收率分别为99.6%-108.7%和99.2%-101.3%,变异系数分别为0.56%-1.08%和0A1%-0.82%。结果表明该方法准确、灵敏度高、简便快速。The simultaneous determination of potassium (K) and sodium (Na) content in urine was established using the atomic absorption spectrometer. The optimized working conditions were ascertained. In order to remove the disturbance of ionization, cesium chloride was added. The disturbance of urinary substrate could be avoided after the urines were diluted to 1:100. The detection limit of K and Na was 0.004 9 mg/L and 0.004 0 mg/L, determination range was 0.2-5.0 mg/L and 0.2-3.0 mg/L, the recovery was 99.6%-108.7% and 99.2%-101.3%, the RSD was 0.56%-1.08%, and 0.41%-0.82%, respectively. The results showed that the method was accurate, sensitive and simple.

关 键 词:火焰原子吸收光谱法   尿液 

分 类 号:O657.31[理学—分析化学]

 