离子色谱同时测定水中10种阴离子含量  被引量:19

Simultaneous determination of 10 negtive ions in water by ion chromatography

在线阅读下载全文

作  者:李文杰[1] 张国棋[1] 

机构地区:[1]辽宁省疾病预防控制中心,沈阳110005

出  处:《中国卫生检验杂志》2008年第11期2285-2287,共3页Chinese Journal of Health Laboratory Technology

摘  要:目的:建立一种离子色谱检测方法,用于同时测定10种阴离子含量。方法:选用Ionpac AS23分离柱,Ionpac AG23保护柱,淋洗液为NaCO3/NaHCO3,浓度为7.0mmol/L/0.4mmol/L。流速1.0ml/min。结果:在0.005~8.00mg/L浓度范围内,各待测物的线性关系良好。结论:该方法简便、灵敏。一次进样23min内,全部分离完全,可满足国家饮用水卫生标准及美国EPA的标准。Objective:To develop a method for simultaneous determination of 10 negtive ions in the water by ion chromatography. Methods:Ten negtive ions were separated by Ionpae AS23 column with Ionpac AG 23 with an eluent containing 7. 0 mmol/ L NaCO3 and 0.4 mmol/L NaHCO3. The flow rate was 1.0 ml/min. Results:The linear concentration range was 0. 005 - 8. 00 mg/L for each agent. Conclusion:The method enables efficient, sensitive and rapid analysis for both standards of EPA and the national drinking water sanitation, which simultaneously and completely separates all agents only in 23 minutes.

关 键 词:10种阴离子 离子色谱 饮用水 

分 类 号:O657.75[理学—分析化学]

 