无银催化-分光光度法测定COD的研究  被引量:4

Research on determination of COD with silver-free catalyst by spectrophometry

在线阅读下载全文

作  者:衣雪松[1] 孙力平[1] 路远[1] 吕建波[1] 袁忠伟[1] 李志伟[1] 

机构地区:[1]天津城市建设学院环境与市政工程系,天津300384

出  处:《环境工程学报》2009年第2期298-300,共3页Chinese Journal of Environmental Engineering

摘  要:以XS-LP为催化剂,探讨了改进的无银催化HACH法测定污水中COD的系列条件实验。结果表明:在药剂A,B的投配比为1∶2,投加总量为12 g/L的条件下,其准确度、精密度与国标法有较好的可比性,分别小于10%和3%;对实际废水的测量结果与国标回流法的相关系数分别高达0.9951和0.9874,可满足一般环境检测技术规范要求。The content of COD in wastewater was determined by using the method of amended HACH which without silver but with XS-LP as catalyst. The results showed that this method was comparable to the standard method used in China in terms of accuracy and precision, were less than 10% and 3% separately, at ratio of A and B was 1:2, gross was 12 g/L. For wastewater in nature, the linear correlation coefficient with standard method is about 0. 9951 and 0. 9874, accord with the requirement of the monitoring skill standard.

关 键 词:COD XS—LP催化剂 精密度 准确度 相关性 

分 类 号:X703[环境科学与工程—环境工程]

 