热敏液晶测温标定及误差分析  被引量:5

Calibration and Error Analysis of Temperature Measurement for Thermal Sensitive Liquid Crystal

在线阅读下载全文

作  者:范玮[1] 吕品[1] 

机构地区:[1]北京航空航天大学能源与动力工程学院,北京100083

出  处:《航空发动机》2009年第1期50-52,6,共4页Aeroengine

摘  要:对SPN/R30C20W型热敏液晶的特性进行了实验研究。实验以相机视角、光源强度、液晶厚度等参数作为对象,讨论了3种参数对液晶测温结果的影响。分析结果表明:在视角小于15°时,相机视角对液晶测温的影响可以忽略;光强不同导致由标定曲线计算出的最大温差达到5.5℃;液晶厚度对标定数据有较大的影响。该实验结论在液晶测温领域广泛应用。The experimental investigation of characteristics of the thermal sensitive liquid crystal whose type is SPN/R30C20W was performed. The camera angle, light source intensity and liquid crystal thickness were regarded as parameters in the experiment and the effects of the three paran^eters on the liquid crystal temperature measurement were discussed. The analysis results show that: the effect of camera angle on the liquid crystal temperature measurement can be neglected when the angle is smaller than 15°; the maximal temperature difference calculated by the calibration curve can be as much as 5.5℃ due to the variety of light intensity; the liquid crystal thickness has remarkable influence on the calibration data, The experimental conclusions can be widely used in the field of liquid crystal temperature measurement.

关 键 词:液晶测温 液晶标定 误差分析 

分 类 号:TU359[建筑科学—结构工程] O753.2[理学—晶体学]

 