毛细管气相色谱分流进样技术研究  被引量:1

Study of Split Injection Technique for Capillary Gas Chromatography

在线阅读下载全文

作  者:郭荣波[1] 王涵文[1] 关亚风[1] 

机构地区:[1]中国科学院大连化学物理研究所,大连116012

出  处:《分析化学》1998年第2期192-195,共4页Chinese Journal of Analytical Chemistry

摘  要:研究出一种改进的毛细管分流进样器,设计了一种新型螺旋进样内衬管。使分流进样的重复精度有显著提高,沸点歧视效应减小,色谱峰尾尖锐,使毛细管色谱柱的高分离效能得以充分发挥。对4种常用的进样内衬进行了考察,得到了有重要实用意义的结果,并从理论上解释了与传统观点相悖的实验现象,深化了对分流进样动态过程的理解。An improved design of split injector and an new cyclo-type split sleeve for capil- lary gas chromatography were developed. The sample discrimination and the tailing of peaks were reduced as compared with commercial injectors, giving better sampling reproducibility and improving separation. The performance of four types of sleeves were evaluated and com- pared, and the contradictory results between this technique and typical technique were inter- prated theoreticqlly. The dynamic process of sample evaporation inside the injector was also discussed.

关 键 词:进样器 设计 分流进样 毛细管气相色谱 

分 类 号:O657.710.2[理学—分析化学]

 