静电场采样装置对室内空气中过敏原采集效率的研究  被引量:5

Study on Collecting Efficiency of Electrostatic Field Sampler for Allergens in Indoor Air

在线阅读下载全文

作  者:张惠力[1] 周明浩[2] 甄世祺[2] 陈连生[2] 要茂盛[3] 陈晓东[2] 周连[2] 张琦[2] 

机构地区:[1]东南大学公共卫生学院,劳动与环境卫生专业江苏南京210009 [2]江苏省疾病预防控制中心,江苏南京210009 [3]北京大学环境与工程学院,北京100871

出  处:《环境监测管理与技术》2009年第2期57-59,共3页The Administration and Technique of Environmental Monitoring

基  金:2007年江苏省科技计划基金资助项目(BS2007069)

摘  要:以Biosampler采样器为参照,采用平行重复的实验方法,以电压和流量作为匹配因素结合ELISA、LAL和GLU-CANTELL技术,检测空气中过敏原Der p和Der f及内毒素和多聚糖。结果表明,静电场采样装置对4种生物气溶胶的采集效率要比Biosampler采样器高2倍~8倍,静电场采样法优于撞击式采样法。Der p,Der f,endotoxin and glucan were collected in parallel and repeatedly by electrostatic sampling devices and biosampler and their concentrations were measured by ELISA,LAL and GLUCANTELL methods.The results showed that the sampling efficiency of electrostatic sampling devices were about 2~8 times higher than biosampler's when collecting the airborne allergens and toxins.The effect of electrostatic sampling was observed better than the impactor.The electrostatic field sampling devices needed to be further improved.

关 键 词:静电场采样器 室内空气 过敏原 Derp Derf 内毒素 多聚糖 

分 类 号:X859[环境科学与工程—环境工程]

 