作 者:赵启仁[1] 李美佳[1] 刘洁[1] 宋娜玲[1] 庄湘莲[1] 陈艾[1] 林汉[1]
机构地区:[1]中国医学科学院中国协和医科大学放射医学研究所
出 处:《生物化学与生物物理进展》1998年第1期71-74,共4页Progress In Biochemistry and Biophysics
基 金:国家自然科学基金
摘 要:报告了应用酶放大镧系元素发光法测定碱性磷酸酶的方法.应用5氟水杨酸磷酸酯作为酶底物,并对方法学中的多种因素进行了最佳比.用甲基硅油(I)改进了信/噪比的稳定性.方法的灵敏度为4U/L.精密度为10%.精密度为10%的浓度范围是200~300×102U/L.测量值的相对误差<10%.测定了血清中的碱性磷酸酶浓度,回收率为93%~95%.An enzymeamplified lanthanide luminescence assay of alkaline phosphatase was described. Optimization of important factors in this assay was studied. Improvement of stability of signal/noise ratio by using methyl silicon oil (I) was first reported. The sensitivity and the accuracy of the assay were 4 U/L and <10% respectively. The concentration range corresponding to 10% precision was 200~302×102 U/L. The activity of alkaline phosphatase in serum was also detected and the recovery was 93%~95%. 5fluorosalicyl phosphate was synthesized and used as substrate.
正在载入数据...
正在载入数据...