双探头符合线路SPECT正电子显像衰减校正实验研究  被引量:4

Phantom attenuation research of 2 detectors SPECT with coincident

在线阅读下载全文

作  者:杜建强[1] 程木华[2] 李小华[3] 朱丽娟[1] 常钢[1] 

机构地区:[1]广州中医药大学附属第二医院、广东省中医院核医学科,510120 [2]中山大学附属第三医院核医学科 [3]广州军区广州总医院核医学科

出  处:《中华核医学杂志》2009年第3期201-204,共4页Chinese Journal of Nuclear Medicine

摘  要:目的用Jaszczak模型及心肌模型检测及评价SPECT和正电子符合探测显像衰减校正在肿瘤和心肌显像中的应用效果。方法用Jaszczak模型充满^18F-脱氧葡萄糖(FDG)溶液,充分混匀,分别在Vertex Plus机型和Hawkeye机型上做模型图像的采集并进行图像重建。结果Vertex Plus衰减校正后的图像质量明显改善,周边和深部的计数之差减少,深部的计数得到补偿。校正后的图像可以发现直径为9mm的“冷”区,而7mm的“冷”区则难以辨认,“热”区可以分辨到6mm。Hawkeye衰减校正后的圆形“冷”区有较大的改善,校正前对直径在12mm以下的“冷”区都无法显示,中央的“冷”区几乎与本底相似,难以分辨,周边的放射性与中央部分的放射性计数差距较大。校正后的图像可以发现5个“冷”区,而7mm的“冷”区则难以辨认。结论衰减校正的模型实验显示用^137Cs作为穿透射线方法的校正效果优于Hawkeye X线球管,不仅明显改变图像质量,而且可以明显改善空间分辨率,在“冷”区可以显示到直径9mm,而“热”区则可以显示到6mm。Objective The photons from the body would be attenuated in their intensity by the soft tissue before they reach the detectors. Some of them lost more intensity than others when coming from the deeper body. Attenuation of the photons would distort the image, affect the image quality, and may lead to misdiagnosis. In this research the X-ray and γ-ray were used as transmission resources for accurate assessment of attenuation in each part of the body. The transmission data was used to correct the emission data from the body. Iteration algorithm was used for the reconstruction of attenuation corrected tomography images. Evaluate the efficacy of attenuation correction in duel head hybrid SPECT/coincidence PET systems (SPECT/PET) and SPECT using both the Jaszczak and the myocardial phantoms. Methods For PET phantom study, the Jaszczak phantom was filled with the aqueous solution of ^18F-fluorodeoxyglucose (FDG) and was scanned with two hybrid SPECT/coincidence PET systems respectively (Vertex Plus, Netherlands Philips ADAC Company and Hawkeye, USA GE Company). The PET images were acquired and reconstructed. Results The "cold" lesion, which the diameter was less than 12 mm of uncorrected image, could not be identified. But it, after correction, diameter 9 mm, would be recognized clearly, less than 7 mm would not be distinguished. Noncorrected image could only identify "hot" lesions which the diameter were 9, 12, 14, 18, 22 and 38 mm, while in the corrected image, not only it was much better, but also the "hot" lesion of 6 mm was distinguished. In Hawkeye, the same situation, the corrected image of was improved obviously ; the disparity of counts between surrounding and deep center less, and the counts in deep center were compensated. There were only 3 round " cold" lesions could be recognized, the diameter: 14, 18 and 20 mm; all of them which diameter were less than 12 mm could not be distinguished. One in the center was almost as background, and could not be identified. The radioactive acc

关 键 词:体层摄影术 发射型计算机 单光子 符合线路 质量控制 

分 类 号:R686[医药卫生—骨科学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象