电感耦合等离子体原子发射光谱法测定易切钢中铅含量  被引量:1

ICP-AES Determination of Pb in Free-cutting Steel

在线阅读下载全文

作  者:王蔚育 

机构地区:[1]上海宝钢建筑工程设计研究院钢结构检测站,上海200050

出  处:《理化检验(化学分册)》2009年第6期724-725,共2页Physical Testing and Chemical Analysis(Part B:Chemical Analysis)

摘  要:采用电感耦合等离子体原子发射光谱法对易切钢中铅的含量进行测定,样品经硝酸-盐酸(1+1)混合酸溶解,选择220.353nm波长为测定铅的分析线,铅的质量浓度在2.00~20.00mg·L^-1范围内与其谱线强度呈线性关系,测定下限(10S/N)为0.03mg·L^-1。该方法用于两个标准样品(GSBH40064—93和GSBH40061—93)中铅含量的测定,加标回收率在97.1%~99.1%之间,相对标准偏差(n=11)小于0.8%。ICP-AES was applied to the determination of lead in free-cutting steel. The sample was dissolved with mixed acid of HNOa and HCI(1+1). Wavelength of 220. 353 nm was selected as analytical spectral line for determination of Pb in sample solution. Linear relationship between values of spectral line intensity and mass concentration of Pb was obtained in the range of 2. 00--20.00 mg·L^-1, with lower limit of determination (10S/N) of 0. 03 mg·L^-1. Standard samples (GSBH 40064-93 and GSBH 40061-93) were analyzed by the proposed method, values of recovery found by standard addition method were in the range of 97. 1 % --99. 1 %, with values of RSDrs (n=ll) less than 0. 8%.

关 键 词:电感耦合等离子体原子发射光谱法 易切钢  

分 类 号:O657.31[理学—分析化学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象