阅片室环境照度和观片灯照度的检测与分析  

Test and Analysis of Illuminance in Reading Room and Viewbox in Department of Radiology

在线阅读下载全文

作  者:杨立娟[1] 张涛[2] 齐乃新[3] 刘波[3] 姚晓新[1] 王俊明[1] 

机构地区:[1]西安市第四医院CT室,陕西西安710004 [2]西北大学信息学院 [3]西安交通大学医学院第二附属医院影像中心

出  处:《实用放射学杂志》2009年第7期1041-1043,共3页Journal of Practical Radiology

基  金:陕西省科技计划资助项目[2006K09-G1(2)];函安市科技攻关资助项目(YF2007138).

摘  要:目的参照有关标准,对影响影像诊断质量因素的阅片室环境照度(IRR)和观片灯照度(IVB)进行检测与分析。方法采用深圳产SAMPO lx1330B数字化照度计对西安市三级医院中38个影像科阅片室及98个观片灯进行IRR、IVB检测,并同时对147个影像诊断医师进行IRR及IVB知识调查。结果(1)各医院均存在IRR过高(不合格率31.6%)、IVB过低(不合格率24.5%)的情况,两者差值〈1900 lx的占被测试总数的42.1%。(2)诊断医师普遍认同应该有合适的阅片室环境照度及观片灯照度,90.3%被调查者认为IRR应低于100 lx,IVB应高于2000 lx。结论西安市各医院阅片室环境照度及观片灯照度不合格率偏高。低IRR及高IVB符合现代医学影像识读的要求。Objective To investigate the factors including illuminance in reading room and viewbox(IRR and IVB) that influence the quality of photographic diagnosis according to the relevant international standard. Methods The tests were made on IRR of 38 radiological reading rooms and IVB of 98 viewboxes in the hospitals of Xi'an with Illuminance - meter of SAMPO lx1330B made in Shenzhen. The knowledges about IRR and IVB were investigated to 147 radiologists. Results (1) IRR in all of the 38 hospitals was generally higher, and 31.6% of it did not come up to the standard. 24.5% of all IVB was lower than the standard. 42. 1% of difference from IVB and IRR was lower than 1900 lx. ( 2 ) Radiologists generally considered that there should he moderate illuminance in reading room and viewbox, 90.3% of them concluded that IRR should be lower than 100 lx and IVB should be higher than 2000 lx. Conclusion IRR and IVB in the hospitals of Xi'an had a high fraction of defect. Increasing IRR and decreasing IVB can meet the photographic diagnosis demand of modern medical imaging.

关 键 词:照度 影像识读 X线摄影 

分 类 号:R814.49[医药卫生—影像医学与核医学] R814.43[医药卫生—放射医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象