检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:张爱丽[1] 项学敏[1] 周集体[1] 曹同川[1]
机构地区:[1]大连理工大学
出 处:《大连铁道学院学报》1998年第2期64-67,共4页Journal of Dalian Railway Institute
摘 要:通过对差压式直读BOD测定仪的改造和操作方法的改善,使仪器的灵敏度提高到改造前的六倍多,最大灵敏度可达25mm/(mg·L-1);同时仪器的精密度也有很大提高,在最大灵敏度下同一水样的平行测试变异系数仅为±3.1%;不仅能测定高BOD值的水样,而且对BOD值小于3mg/L的样品也能得以可靠测量,扩大了该仪器的使用范围,提高了仪器的使用效率.The sensitivity of the BOD measure instrument with different pressure and directly reading is increased by more than six times through both reforming some parts of the instrument and improving some mehods of operation.The greatest sensitivity can be 25 mm/(mg·L -1 ).Meanwhile,the precision of the instrument becomes much better than ever before.At the greatest sensitivity the coefficient of variation is only ±3.1% for the same sample of water.After being reformed,the instrument can not only be used to measure the high BOD sample of water but also be used to measure reliably the low BOD sample of water,especially less than 3 mg/L Therefore the measuring uange of the instrument of extended and the using efficiency of the instrument is raised.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.129.39.144