血糖仪偏差图设计与临床应用  被引量:1

Bias Plots of Glucose Meter Systems And Clinical Applications

在线阅读下载全文

作  者:夏寿扬[1] 史菊[1] 王娟[1] 

机构地区:[1]江苏省泗洪县人民医院检验科,江苏泗洪223900

出  处:《现代检验医学杂志》2009年第4期123-124,共2页Journal of Modern Laboratory Medicine

摘  要:目的探讨血糖仪准确性评价方法。方法以血糖仪和参比法结果之差绘制血糖仪偏差图,判断血糖仪的准确性。结果依据IS015197准则,当血糖≤75mg/dI(4.2mmol/L)时,三种血糖仪符合率分别为73.53%,97.06%和100%,〉75mg/dl(4.2mmol/L),符合率分别为89.16%,99.40%,99.40%。结论评价血糖仪准确性应当遵循ISO15197准则,“血糖仪偏差图”能够直观地显示血糖仪结果的准确性。Objective To study glucose meter systems accuracy. Methods "Bias plots" were done by differeces value of all paired glucose meter systems and reference measurements and meters were evaluted. Results Depending on ISO15197 guideline,according rates of three meters were 73. 53%,97.06% and 100% respectively when results ≤75 mg/d,while〉 75 mg/dl were 89. 16% ,99.40% and 99.40%. Conclusion Evaluation of meters must follow ISO15197 guideline. Accuracy of systems can be judged by "bias plots"visually.

关 键 词:血糖仪偏差图dS015197准则 

分 类 号:R446[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象