摘 要:采用离子交换固定法把Luminol(Lu)、CN-固定在阴离子交换树脂上,串入流动注射系统中。当一定量洗脱剂被注入阴离子交换柱时,洗脱下来的Lu和CN-与分析物Cu2+发生化学反应,产生化学发光,实现对Cu2+的在线检测。本法检出限为1.3×10-9g/mL,线性范围为4.0×10-9~6.0×10-6g/mL,RSD为3.4%,单次测定在1min内能完成。已用于井水中铜的测定,结果满意。Abstrac A new chemiluminescence flow system has been developed in this paper.The analytical reagents luminol and CN - were both immobilized on an anion exchange resin column.When sodium phosphate was injected into the anion exchange resin column,luminol and CN - were eluted ,and then reacted with Cu 2+ to produce chemiluminescence,by which Cu 2+ was detected.The concentration range of linear response is from 4.0×10 -9 to 6.0×10 -6 g/mL,and the detection limit is 1.3×10 -9 g/mL.The method has been used for the determination of copper in well water with satisfactory results.