猪冠状动脉内金属支架蛋白涂层的生物相容性研究  被引量:21

The Biocompatibility of Protein Coating Applied to Metallic Stent

在线阅读下载全文

作  者:袁晋青[1,2] 高润霖[1,2] 史瑞文[1,2] 宋来凤 汤健[1,2] 李永利[1,2] 唐承君 孟亮[1,2] 袁卫民 陈在嘉[1,2] 

机构地区:[1]中国医学科学院 [2]中国协和医科大学

出  处:《中国循环杂志》1998年第5期271-271,共1页Chinese Circulation Journal

基  金:国家"九五"攻关课题

摘  要:目的:评估蛋白涂层金属支架植入冠状动脉的生物相容性和致血栓性,并通过造影及病理学检查以确定其对动脉壁的影响。方法:金属支架由316L医用不锈钢丝编织而成,将该支架浸入含有交联剂的明胶溶液一定时间后,经干燥制成蛋白涂层。将蛋白涂层支架固定在3.0mm经皮冠状动脉腔内成形术(PTCA)球囊上,通过8F大腔引导导管植入13只小型猪冠状动脉前降支中段。结果:11只动物支架被成功植入前降支中段,2只动物支架植入前死于心室颤动。支架植入后3天及7天造影证实管腔通畅,无血栓形成、充盈缺损、夹层、栓塞、支架移位或血管痉挛。扫描电镜显示3天时有少量血小板沉积和纤维蛋白形成,且部分内皮化。7天时完全内皮化。4只动物在28天造影时管腔通畅。支架植入节段的所有管腔狭窄直径均<50.0%,平均狭窄直径为(30.0±7.2)%,无闭塞发生。组织病理学检查显示:所有支架植入节段完全内皮化,有不同程度的新生内膜增生,无异物巨细胞聚集,局部管壁完整,无中层坏死。平均新生内膜厚度为0.33±0.17mm,平均新生内膜面积为0.41±0.31mm2,平均百分狭窄面积为(32.0±21.0)%。结论:在小型猪冠状动脉模型中,通过球囊导管系统植入蛋?Objective:To assess the biocompatibility and thrombogenicity of crosslinked protein coating of the metallic stent. Methods:Metallic stent was woven by 316 L medical stainless steel.The coating of metallic stent was made by immersing the stent in gelatin solution containing crosslinker and then air dried.The coated stents mounted on 3.0 mm percutaneous transluminal coronary angioplasty balloon and implanted in the left anterior descending (LAD) of coronary arteries through 8F guiding catheter in 13 mini swine. Results:The stents were successfully implanted into the middle segment of LAD in 11 animals.Two animals died of ventricular fibrillation before stenting.Angiography revealed that the vessels were patent 3( n =3)and 7( n =4)days after stent implantation,without evidence of thrombosis,filling defects,dissection,embolization,stent migration or vessel spasm.At the 3rd day,scanning electron microscope and histopathologic examination showed minimal platelet deposition,fibrin formation and partial endothelialization on the stent,at the 7th day complete endothelialization developed.At the 28th day coronary angiography( n =4) demonstrated vessel patency and diameter stenosis <50.0%[mean diameter stenosis was (30.0±7.2)%],without occlusion,histopathologic examination revealed complete endothelialization at the site of stent implantation and a variable degree of neointimal proliferation.There were no large collection of foreign body giant cells,no medial necrosis,local vascular integrity was maintained. Mean neointimal thickness and neointimal areas were 0.33±0.17 mm and 0.41±0.31mm 2,respectively.Mean percentage area stenosis was (32.0±21.0)%. Conclusion:In the mini swine model,protein coated stent can be easily implanted into coronary artery by balloon catheter. There was no serious local vascular tissue reaction.The coating of the stent has good biocompatibility without thrombogenesis.

关 键 词:支架 蛋白涂层 生物相容性 血管成形术 

分 类 号:R654.3[医药卫生—外科学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象