流动注射化学发光法测定多巴胺  被引量:5

DETERMINATION OF DOPAMINE BY FLOW INJECTION ANALYSISWITH CHEMILUMINESCENCE DETECTION

在线阅读下载全文

作  者:李丽清[1,2,3] 孟兆珂[1,2,3] 吕九如[1,2,3] 

机构地区:[1]泰安师范专科学校化学系 [2]泰安市肿瘤医院 [3]陕西师范大学化学系

出  处:《泰山医学院学报》1998年第2期115-117,共3页Journal of Taishan Medical College

基  金:国家自然科学基金

摘  要:目的建立化学发光测定多巴胺新方法。方法采用流动注射技术,研究了多巴胺———高锰酸钾体系的化学发光行为。结果方法的检出限为6×109g/ml,线性范围为20×108~80×105g/ml。对40×107g/ml多巴胺进行11次平行测定,相对标准偏差为18%。结论该方法用于药剂中多巴胺含量测定,结果与药典标准方法测得值一致。Objective To establish a new chemiluminescence method of determinating dopamine. Methods: The flow injection technique was used, and the chemilumines cence reaction of dopamine-KMnO4 was studied. Results Detection limit was 6×109g/ml. The linear range was 2.0×108~8.0×105g/ml. The parallel determination was done 11 times to 4.0×109g/ml dopamine and the relative standard deviation was 1.8%. Conclusion Chmiluminescence with flow injection is in accordance with standard method in the determination of dopamine.

关 键 词:多巴胺 化学发光 流动注射 

分 类 号:R971.93[医药卫生—药品]

 