作 者:刘婧[1] 周亦伦[1] 孙芳[1] 马丽洁[1] 沈洋[1] 韩彬[1] 刘文虎[2] 崔太根[1]
机构地区:[1]首都医科大学附属北京朝阳医院,北京100020 [2]首都医科大学附属北京友谊医院肾内科,北京100050
出 处:《中国血液净化》2010年第2期84-87,共4页Chinese Journal of Blood Purification
摘 要:目的探索应用改良的节段生物电阻抗法-小腿电阻抗比值(CBIR)评估血液透析(HD)患者的容量状况。方法使用多频生物电阻抗分析仪测定157例健康志愿者右侧小腿的电阻抗值,计算CBIR(impedance 200k/impedance 5k)作为容量负荷指标,并确定正常参考值范围。同时检测117例HD患者透析后CBIR,观察患者容量负荷状况。结果健康志愿者的CBIR为(0.774±0.028),由此确定正常参考值范围为0.746~0.802(均数±1个标准差)。HD组透析后CBIR明显高于正常对照组[(0.790±0.033)vs(0.774±0.028),P<0.001],其中容量增高者(CBIR>0.802)、容量正常者(CBIR在0.746~0.802)和容量降低者(CBIR<0.746)分别占36.8%,53.0%,10.2%。在使用降压药的剂量无显著差异的情况下,CBIR增高组高血压的发生率明显高于CBIR正常组(P=0.008)和CBIR降低组(P=0.003)。结论小腿电阻抗检测可能是评估HD患者容量状况的有效方法。Objective To evaluate volume status in hemodialysis (HD) patients by modified bioimpedance spectroscopy - calfbioimpedance ratio (CBIR). Methods Calfbioimpedance was measured by bioimpedance spectroscopy. As an index of volume status, CBIR was calculated as follows: CBIR = impedance of 200 kHz / impedance of 5kHz. Reference CBIR range was obtained from 157 healthy volunteers. The volume status of 117 HD patients post-dialysis was assessed using the CBIR from healthy volunteers as the reference CBIR range. Results CBIR of the healthy volunteers was 0.774 + 0.028, the reference range was thus defined as 0.746-0.802 (mean ± SD). CBIR was significantly higher in patients post-hemodialysis than in healthy volunteers (0.790 ± 0.033 vs 0.774 ±0.028, P 〈0.001). HD patients were divided into three groups according to the CBIR: high CB1R group (〉0.802), normal CBIR group (0.746 0.802), and low CBIR group (〈 0.746). There were 43 patients in high CBIR group (36.8%), 62 patients in normal CBIR group (53.0%), and 12 patients in low CBIR group (10.2%). The doses of antihypertensive agents were not significantly different among the three groups. Nevertheless, hypertension was significantly higher in high CBIR group than in normal CBIR group (p=0.008) and low CBIR group (p=0.003). Conclusions Calf bioimpedance measurement may be an effective method for the evaluation of volume status in hemodialysis patients.
分 类 号:R318.16[医药卫生—生物医学工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.244