检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:周云龙[1,2] 吴粦华[1] 胡志彪[1] 拓宏桂[1]
机构地区:[1]龙岩学院化学与材料系,龙岩364000 [2]福州大学化学化工学院,福州350000
出 处:《理化检验(化学分册)》2010年第3期235-237,共3页Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
基 金:褔建省教育厅自然科学科研课题(JA03165)
摘 要:用线性扫描极谱法研究盐酸二甲双胍的电化学行为。在pH6.85的0.1mol·L-1磷酸氢二钠-磷酸二氢钾缓冲溶液中,盐酸二甲双胍于-1.165V(vs.SCE)处产生一灵敏的吸附波,其一次微分线性扫描峰电流与盐酸二甲双胍质量浓度在0.41~41.0mg·mL-1范围内呈线性关系,检出限(3S/N)为0.28mg·mL-1。对41.0mg·mL-1盐酸二甲双胍溶液进行6次平行试验,相对标准偏差(n=6)为0.72%。此方法已用于测定其片剂中盐酸二甲双胍的含量。It was found that in a PBS of pH 6. 85 and by the linear scanning polarography, a sensitive adsorption wave of metformin was observed at the potential of -1. 165 V (vs. SCE). Linear relationship between values of the 1st derivative of values of peak current and mass concentration of metformin was obtained in the range of 0. 41--41.0 mg ~ L-x with detection limit (3S/N) of 0. 28mg·L^-1. Precision was tested with standard solution of metformin hydrochloride of 41.0 mg· L^-1 for 6 determinations, value of RSD found was 0. 72%. The proposed method has been used in the determination of content of metformin hydrochloride in its tablets.