压电双晶片在眼压测量中的实验研究  

The Experiment and Research for Piezoelectric Bimorphs in Measuring Intraocular Pressure

在线阅读下载全文

作  者:田学隆[1] 向逾[1] 

机构地区:[1]重庆大学生物工程学院,重庆400044

出  处:《压电与声光》2010年第2期243-246,共4页Piezoelectrics & Acoustooptics

基  金:重庆市"十一五"科技重大专项"新型医疗器械"基金资助项目(CSTC2008AA5010)

摘  要:介绍了一种利用压电双晶片的逆压电效应来测量眼压的方法及其装置。该方法建立了流过晶片的电流和眼球内部压强的关系,通过检测的电流值来测量眼压值。实验表明,该方法操作简便,可行性较好,有一定的临床应用前景。该文还给出了实验结果和误差分析。A method taking the effect of the anti-piezoelectricity of the piezoelectric bimorphs to measure intraocular pressure was proposed. This method established relationships between current washing through the bimorphs and intraocular pressure. Then the result of intraocular pressure was valued by the measured current. The experiment showed that it was practicable and simple for operation and with some application prospect in clinic. Experiment results and error analysis were also given in the paper.

关 键 词:压电双晶片 振动 眼压 曲线拟合 