检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:赵洪斌[1,2] 张新[1] 包尚联[1,3] 黄斐增[1] 李泉凤[4]
机构地区:[1]北京大学医学物理和工程北京市重点实验室 [2]北京医疗器械研究所研发部 [3]北京海思威科技有限公司研发部 [4]清华大学理学院
出 处:《Transactions of Nanjing University of Aeronautics and Astronautics》2010年第1期7-12,共6页南京航空航天大学学报(英文版)
基 金:Supported by the National Natural Science Foundation of China (60672104,10675013);the Na-tional Basic Research Program of China ("973"Program)(2006CB705705);the 10th Five-Year Plan of the Ministry of Science and Technology of China(2001BA706B-05);the Joint Research Foundation of Beijing Municipal Commissionof Education~~
摘 要:A method for designing an X-ray flatness filter for medical electron linac is developed. It is used in the optimization process in the electron beam radiation system. Monte Carlo simulation method is used and two examples of real radiation system optimization processes for China-made medical electron linac are provided: 15 MV X- ray system of BJ-20 linac, and 12 MeV electron system of BJ-14. Results are verified by using the traditional method.提供了一种用于设计医用电子直线加速器(linac) X射线均整过滤器的方法,并将其用于电子束辐射系统的优化过程之中。采用Monte Carlo模拟计算方法,提供了2个真实的国产医用电子直线加速器的辐射研究成果:BJ-20加速器的15 MV X射线辐射系统和BJ-14加速器的12 MeV电子束辐射系统。使用传统方法对上述成果进行了验证。
关 键 词:linear accelerators computer simulation and Monte Carlo methods radiation system optimization flatness filter design