机构地区:[1]航天医学工程研究所 [2]国防科技大学
出 处:《航天医学与医学工程》1999年第1期51-55,共5页Space Medicine & Medical Engineering
摘 要:目的对应用双频率阻抗测试和自适应处理技术相结合的方法消除呼吸阻抗测量中的运动干扰问题进行一些探索性的研究。方法采用自行研制的双通道呼吸阻抗测量硬件系统和以自适应抵消处理为主的软件系统进行了人体实验研究。结果运动干扰信号基本消除。低频与高频呼吸阻抗信号的比值K随着运动强度的增加而增大。结论应用该方法消除呼吸阻抗测量中的运动干扰具有一定的可行性,但是在实际应用中,为了得到更好的结果,应注意考虑K值的准确取值。Objective To study the application of dual frequency impedance pneumography and adaptive filter for reducing motion artifacts in respiration monitoring.Methods Human experiments were done by using a dual-channel plethysmograph and the software including two adaptive processor.Results It was found that the motion artifacts were essentially reduced and the amplitude ratio(K) of low to high frequency signals of respiration increased with the intensity of movement.Conclusion The dual frequency pneumography and adaptive filter is practicable in respiration monitoring for reducing motion artifacts.But K should be rectified according to conditions so as to obtain better results of artifact reduction.