在用血压计质量状况分析  被引量:1

Analysis on Quality of Sphygmomanometers on Service

在线阅读下载全文

作  者:王慧梅[1] 谢华[1] 刘光荣[2] 吴昊[2] 

机构地区:[1]新疆军区药品仪器检验所,乌鲁木齐830063 [2]解放军总医院医学工程保障中心,北京100853

出  处:《医疗卫生装备》2010年第7期90-90,99,共2页Chinese Medical Equipment Journal

摘  要:调查研究了新疆军区驻乌鲁木齐市部队在用血压计、血压表的使用状况。依据《JJG270—2008血压计和血压表检定规程》对该部队的605台血压计、血压表进行检定,结果表明示值误差超差、气密性不符合要求以及袖带破损是造成血压计、血压表不合格的主要因素,为血压计、血压表的正确使用和质量控制提供了借鉴。The use state of sphygmomanometers on service in troops quartered at Urumchi of Sinkiang Military Area command is investigated.Also 605 sphygmomanometers are metrologically verified according to the verification regulations for sphygmomanometers JJG270—2008;the results demonstrate that indication error,non-conformance of tightness and breakage of cuff are the main factors causing disqualification of the sphygmomanometers and these could be referenced as the useful experience for guiding the appropriate using and quality control of sphygmomanometers.

关 键 词:血压计 计量检定 质量分析 

分 类 号:TH776[机械工程—仪器科学与技术] TB9[机械工程—精密仪器及机械]

 