作 者:李国萍[1] 南一权[1] 高广慧 慕善学[3] 陈慧君[1]
机构地区:[1]辽河石油勘探局妇婴医院,辽宁盘锦124010 [2]辽宁省药品检验所,辽宁沈阳110023 [3]沈阳药科大学高等职业技术学院,辽宁沈阳110026
出 处:《辽宁中医药大学学报》2010年第8期193-194,共2页Journal of Liaoning University of Traditional Chinese Medicine
摘 要:目的:建立α-生育酚残留溶剂的检查方法。方法:用顶空气相色谱法对α-生育酚中乙醇、乙酸乙酯、苯残留溶剂进行检测,采用rtx-5(5%苯基-95.0%二甲基聚氧硅烷)(30m×0.32μm×0.32mm),程序升温,顶空进样,进样体积1mL。结果:精密度RSD均小于2%,峰面积与浓度均有良好的线性关系,最低检测限在67ng·mL-1以下,回收率均符合要求。结论:该方法稳定、简便、准确,可用于α-生育酚3种有机溶剂残留物检测。Objective: To develop a GC method for analysis of residual organic solvents(ethanol,ethyl acetate and benzene) in α-tocopherol. Methods:The contents of residual organic solvents(ethanol,ethyl acetate and benzene) in α-tocopherol were determined by headspace GC on rtx-5 column(30m×0.32μm ×0.32mm),which was packed by 5% benzyl-95.0% dimethyl polysiloxane,the temperature was raised in program and the volume of injection was 1mL. Results:There was good linearity between concentration and peak area. The RSD of precision were less than 2%,the lowest detection limits were below 67ng·mL-1 and the results of recoveries were satisfied with the requirements. Conclusion:This method could be used in determining the residual organic solvents(ethanol,ethyl acetate and benzene) inα-tocopherol,which was stable,convinient and accurable.