全自动生化分析仪分析项目间试剂交叉污染及避免方法  被引量:7

The methods to avoid cross-contamination in automatic biochemical analyzer

在线阅读下载全文

作  者:葛鑫[1] 刘刚[2] 

机构地区:[1]蚌埠医学院第一附属医院检验科,安徽蚌埠233004 [2]安徽省蚌埠市第三人民医院检验科,233000

出  处:《蚌埠医学院学报》2010年第9期943-944,共2页Journal of Bengbu Medical College

摘  要:目的:探讨全自动生化分析仪分析项目间试剂交叉污染及避免的对策。方法:用质控血清分别先测血清碱性磷酸酶(ALP)后测Mg2+(A组);先测淀粉酶(AMY)后测Mg2+(B组);用质控血清测定Mg2+(对照组)。用去离子水当样品测1次ALP后,再连测10次Mg2+,用质控血清替代去离子水当样品同样操作(C组);将ALP换成AMY进行同样操作(D组)。结果:A、B两组污染率差为9.93%和10.13%,ALP和AMY试剂对Mg2+测定有交叉污染。结论:检验人员熟悉仪器的操作及性能,在仪器项目顺序安排时尽量将有污染的项目隔开,按一定的顺序安排项目,可以避免试剂间的化学污染。Objective:To discuss the treatment of avoiding the different item′s cross-contamination of automatic biochemical analyzer.Methods:There were divided into five groups: group A: the contents of alkaline phosphatase(ALP) in patients′ serum with quality control serum were verified,then Mg2+(magnesia ion) was verified;group B:firstly amylase(AMY) then Mg2+ were verified;matched group: just assess Mg2+;group C: the deionizer water was considered as sample,firstly assess the content of ALP in deionizer water,then assess Mg2+ ten times,afterward do the same processes which using quality control serum instead of deionizer water;group D:Use AMY instead of ALP,and do above-mentioned procedure.Results:Group A and group B the pollution rate were 9.93% and 10.13%.The cross-contamination was happened in assessment of Mg2+ as affected by reagent ALP and AMY.Conclusions:Medical technologists must be familiar with the manipulation and function of machines.When arranging the order of machines′ testing items,it is needed to separate these items which has cross-contamination as far as possible.With the specific order,chemical pollution in different reagents can be reduced.

关 键 词:诊断 实验室 碱性磷酸酶 淀粉酶 

分 类 号:R446[医药卫生—诊断学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象