Infinia Hawkeye CT球管维护  被引量:7

Maintenance of Computed Tomography Tube of Infinia Hawkeye CT

在线阅读下载全文

作  者:刘金华[1] 欧阳伟[1] 冯会娟[1] 胡瑞[1] 刘伟英[1] 黄柳华[1] 

机构地区:[1]南方医科大学珠江医院核医学科,广州510282

出  处:《医疗卫生装备》2010年第10期120-121,共2页Chinese Medical Equipment Journal

摘  要:目的:研究Infinia Hawkeye CT球管在不同室温下的温度变化情况。方法:调整机房温度,观察CT球管附近温度变化。结果:机器连续运转,CT球管附近温度逐步升高,机房温度调高,CT球管附近温度会更高。结论:机房温度为18℃时,CT球管附近温度变化最小。Objective To investigate the temperature change of the computed tomography tube of Infinia + Hawkeye in different indoor temperatures.Methods The temperature changes near the computed tomography tube is investigated by coordinating the temperature in the Emission Computed Tomography machine room.Results The temperature near the computed tomography tube rises gradually with the running of the machine,which gets higher as the temperature in the Emission Computed Tomography machine room increases.Conclusion Temperatures near the computed tomography tube changes the lest when it is 18℃ in the room.

关 键 词:Infinia Hawkeye CT球管 温度 维护 

分 类 号:TH774[机械工程—仪器科学与技术] TM407[机械工程—精密仪器及机械]

 