Construction of Tissue-Equivalent Proportional Counter System  

Construction of Tissue-Equivalent Proportional Counter System

在线阅读下载全文

作  者:ZHANG Wei-hua,WANG Zhi-qiang 

出  处:《Annual Report of China Institute of Atomic Energy》2009年第1期252-252,共1页中国原子能科学研究院年报(英文版)

摘  要:In order to establish a set of TEPC system,we tested the characteristic of TEPC in gamma and neutron fields respectively,then calibrated it in the 252Cf source neutron field.The microdosimetryIn order to establish a set of TEPC system, we tested the characteristic of TEPC in gamma and neutron fields respectively, then calibrated it in the ^252Cf source neutron field. The microdosimetry spectra are obtained from the pulse amplitude spectra, and the neutron and gamma components are analyzed. According to the measured spectra (Fig. 1), the absorbed dose rate and dose equivalent rate can be calculated. Simulation calculation was performed by Monte Carlo procedure, whose results are in agreement with the measured data, the values are listed in Table 1. Then, we know the measurement of TEPC system is practicable.

关 键 词:组织等效正比计数器 EPC系统 施工 蒙特卡罗程序 模拟计算 剂量当量率 伽马射线 微剂量谱 

分 类 号:TL506[核科学技术—核技术及应用] TP271.32[自动化与计算机技术—检测技术与自动化装置]

 