检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]f陕西师范大学化学与材料科学学院陕西省生命分析重点实验室,西安710062
出 处:《化学学报》2011年第2期199-203,共5页Acta Chimica Sinica
摘 要:建立了一种核壳型SiO2荧光纳米粒子标记的激光诱导荧光免疫测定的新方法.该方法用Ru(bpy)32+掺杂的核壳型SiO2荧光纳米粒子为标记物,采用双抗体夹心法,在载玻片上固定鼠抗人癌胚抗原(CEA)单克隆抗体,加入1~2μL样品,再加入纳米粒子标记的另一种鼠抗人癌胚抗原(CEA)单克隆抗体.识别后在光导纤维激光诱导荧光毫米阵列检测平台上测量荧光信号,进行定量.在优化条件下,荧光强度和癌胚抗原的量在1~80 pg的范围内具有良好的线性关系,相关系数r=0.9933,方法的检出限为0.3 pg(20 amol).对40 pg的CEA平行测定5次,相对标准偏差为5.5%.A new immunoassay method for Carcinoembryonic Antigen based on fluorescence induced by laser and labeled with nanoparticles is presented.Core-shell SiO2 fluorescent nanoparticles were used as la-bels,"sandwich" model was used.Mouse anti-carcinoembryonic antigen(CEA) monoclonal antibody was immobilized on slides,then 1~2 μL CEA was added,and the other mouse anti-CEA monoclonal antibody labeled with nanoparticles was added,after CEA and antibody recognized,the fluorescent signal was quanti-fied with an laser-induced fluorescence millimeter array detection platform based on optical fiber.The fluo-rescence intensity has a linear relationship with the concentration of CEA in the range of 1~80 pg,with the absolute limit detection of CEA being 0.3 pg.The relative standard deviation(RSD) for 5 parallel measure-ments of CEA(40 pg) was 5.5%.
分 类 号:TB383.1[一般工业技术—材料科学与工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.249