国外原装与国产比色杯在ACL TOP血凝仪应用中的一致性分析  被引量:1

Consistency of Household Colorimetric Cylinder and Original Packing Ones in ACL TOP Automatic Coagulometer

在线阅读下载全文

作  者:梁伟[1] 陈晓蓓[1] 

机构地区:[1]浙江省宁波市第一医院检验科,315010

出  处:《国际检验医学杂志》2011年第2期183-184,共2页International Journal of Laboratory Medicine

摘  要:目的探讨国外原装与国产比色杯在ACL TOP血凝仪应用中的一致性。方法选取24份临床血凝检测标本,分别用原装配套的比色杯和国产比色杯进行检测,对检测结果进行Bland-Altman统计分析。结果 PT、FIB、APTT、TT检测结果在一致性界限外的概率很低,在一致性界限内的最大差值可接受。结论国产和国外原装比色杯一致性良好,在ACL TOP血凝仪中可用国产比色杯取代原装产品。Objective To discuss the difference between household colorimetric cylinder and original packing cups in ACL TOP Automatic coagulometer. Methods 24 clinical samples were chosed for coagulation analysis by household colorimetric cylinder and original packing cups respctviely,and Bland-Altman analysis were applied to analyse these data abtained. Results There is no significant difference were observed out of the limit of consisitance for both the four item including PT,FIB,APTT,TT,and the big- gest difference could be accepted in clinic in the limit of consisitance. Conclusion The household colorimetric cylinder and original packing cups can get the same results and we can use household cylinder to replace the original packing ones.

关 键 词:系统分析 Bland-Altman 一致性评价 

分 类 号:R197.39[医药卫生—卫生事业管理]

 