检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]山东威高集团医用高分子制品股份有限公司,威海264209
出 处:《中国医疗器械信息》2011年第2期34-37,共4页China Medical Device Information
摘 要:分别用TPE输液器和PVC输液器模拟临床输液操作,发现硝酸甘油、尼莫地平和单硝酸异山梨酯注射液经PVC输液器后药液成分有明显丢失现象,而这三种药液经TPE输液器后几乎没有丢失现象。两种材料的输液器对盐酸多巴酚丁胺和呋塞米都没有吸附现象。试验证明,TPE输液器药物兼容性优于PVC输液器。Simulated clinical infusion operation by using separately TPE and PVC infusion set ,It have been discovered that Nitroglycerine, Nimodipine and Isosorbide Mononitrate have the obvious loss phenomenon after flowing the PVC Infusion Set. However, Those drugs has not nearly the loss phenomenon after flowing the TPE Infusion Set. In addition, The dobutamine hydrochloride and furosemide have not been both adsorbed after flowing the TPE and PVC Infusion Set.It is proved that the TPE Infusion Set have an advantage over the PVC Infusion Set at the Medicine compatibility.
关 键 词:药物相容性 TPE输液器 PVC输液器 吸附 紫外吸光度
分 类 号:R197.39[医药卫生—卫生事业管理]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.138.170.222