作 者:胡镜清[1] 赵婷[1] 徐贵成[1] 吴烈[1] 寇秋爱[2] 庄红[2] 李振华[2] 宿明良 张广福
机构地区:[1]中国中医科学院广安门医院,北京100053 [2]中国中医科学院西苑医院,北京100091 [3]北京斯脉福科技发展有限公司,北京100027
出 处:《世界科学技术-中医药现代化》2011年第1期74-77,共4页Modernization of Traditional Chinese Medicine and Materia Medica-World Science and Technology
基 金:国家科技重大专项课题(008ZX09312-012):中药新药临床研究技术平台规范建设;负责人:胡镜清
摘 要:目的:评价SM-1A中医四脉脉诊测量仪脉象诊断的准确性。方法:应用诊断性试验设计,评价SM-1A中医四脉脉诊测量仪对120例受试者溢脉、聚脉、韧脉、紊脉4种基本脉象与中医专家脉诊结果对照的一致性及其自身重复性。结果:与中医专家脉诊结果对照,SM-1A中医四脉脉诊测量仪对溢脉、聚脉、韧脉、紊脉的诊断敏感度分别为0.859、0.945、0.957和0.898;特异度分别为0.875、0.8、0.907和0.833;Kappa值分别为0.733、0.631、0.763和0.549;正确率为0.867,0.933,0.917和0.892;Youden指数分别为0.734、0.745、0.864和0.731。其自身两次重复诊断的完全符合率为81.7%。结论:SM-1A中医四脉脉诊测量仪可以较为准确地进行四脉诊断,具有较好的信度和效度。This study aimed to assess the reproducibility of SM-1A TCM pulse detector according to traditional Chinese medicine (TCM) experts. A total of 120 subjects were enrolled. Each subject was diagnosed by the SM-1A TCM pulse detector and TCM expert, respectively. And the pulse diagnosis was based on four types of pulse, including yimai, ju-mai, ren-mai and wen-mai. The Kappa test was applied. Compared to TCM expert, the sensitivity of the pulse detector for yi-mai was 0.859, for ju-mai was 0.945, for ren-mai was 0.957, and for wen-mai was 0.898. The specificity of the pulse detector for yi-mai was 0.875, for ju-mai was 0.8, for ren-mai was 0.907, and for wen-mai was 0.833. The Kappa value for yi-mai was 0.733, for ju-rnai was 0.631, for ren-mai was 0.763, and for wen-real was 0.549. The coincidence rate for yi-mai was 0.867, forju-mai was 0.933, for ren-mai was 0.917, and for wen-mai was 0.892. The Youden index for yi-mai was 0.734, for ju-mai was 0.745, for ren-mai was 0.864, and for wen-mai was 0.731. The coincidence rate of SM-1A TCM pulse detector in the successive diagnosis of the same patient was 81.7%. Therefore, the SM-1A TCM pulse detector is able to diagnose the pulse accurately based on the four types of pulse.The test-retest reliability was good.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.145.109.97