机构地区:[1]扬州工业职业技术学院,扬州225127 [2]江苏联环药业集团有限公司,扬州225009
出 处:《分析仪器》2011年第2期27-29,共3页Analytical Instrumentation
摘 要:用HPLC法测定了氢化可的松戊酸酯有关物质,结果表明,氢化可的松戊酸酯与氢化可的松及氢化可的松-21-戊酸酯分离良好,氢化可的松戊酸酯与氢化可的松及氢化可的松-21-戊酸酯的最低检测限分别为0.9ng,0.2ng,2.0ng。本法专属性强,结果准确、灵敏,操作简便,可作的质量控制方法。Objective: To establish a HPLC method for the determination of hydrocortisone valerate and related substances.Method: A Phenomenex C18 column(250mm×4.6mm,5μm) was used with ACN-water(45 ︰55) as mobile phase.The flow rate was 0.8mL/min,the column temperature was 30℃,and the detection was at 290nm.Results: Hydrocortisone valerate was separated from impurities(hydrocortisone and hydrocortisone-21-valerate) completely.Their detection limits for hydrocortisone valerate,hydrocortisone and hydrocortisone-21-valerate were 0.9ng,0.2ng and 2.0ng,respectively.Conclusion: The method is accurate,sensitive,simple and specific,and is appropriate for the quality control of hydrocortisone valerate.
关 键 词:高效液相色谱法 氢化可的松戊酸酯 氢化可的松 氢化可的松-21-戊酸酯