不同层厚对128层螺旋CT肺动脉显示率及伪影发生率的影响  被引量:3

Influence of different slice thickness on display rate of pulmonary arterial and incidence rate of artifacts by 128-slice spiral CT

在线阅读下载全文

作  者:刘宁[1] 牛玉军[1] 曹阿丹[1] 李国策[1] 

机构地区:[1]辽宁医学院附属第一医院放射科,辽宁省锦州市121001

出  处:《中国组织工程研究与临床康复》2011年第4期675-678,共4页Journal of Clinical Rehabilitative Tissue Engineering Research

摘  要:背景:行螺旋CT肺动脉造影检查,肺动脉显示率与伪影发生率,对诊断准确性评价有着重要作用。目的:评价不同层厚对128层螺旋CT肺动脉显示率及伪影发生率的影响。方法:经螺旋CT肺动脉造影检查将疑似急性肺栓塞患者图像分别重建为不同层厚的4组:0.6,1.0,3.0,5.0mm,比较4组图像对栓塞肺动脉显示的清晰程度,并评价4组图像显示肺动脉各级分支的连续程度与伪影发生率。结果与结论:主肺动脉、左右肺动脉及叶动脉显示,4组间差异无显著性意义;段及亚段级肺动脉显示,0.6,1.0mm层厚组优于3.0,5.0mm组(P<0.001);第6级肺动脉显示,4组间差异均有显著性意义(P<0.001),且0.6mm组显示率最佳。1.0mm层厚图像在显示亚段级以上肺动脉分支方面与0.6mm层厚能力相近,且优于3.0,5.0mm层厚组;3.0,5.0mm组显示肺动脉各级分支连续程度低于0.6,1.0mm组。39例阴性患者部分容积伪影0.6mm组2例,1.0mm组5例,3.0mm组11例,5.0mm组17例,组间比较差异有显著性意义(P<0.001)。提示,行螺旋CT肺动脉造影检查,采用3.0mm及以下层厚可较准确观察亚段肺动脉;采用0.6mm层厚可减少部分容积伪影,从而提高螺旋CT肺动脉造影图像质量。BACKGROUND:The display rate of pulmonary artery and the incidence of artifact by CT pulmonary angiography (CTPA) play an important role in the evaluation of diagnostic accuracy.OBJECTIVE:To evaluate the effect of different thickness on the 128-slice spiral CT display rate of pulmonary artery and the incidence rate of artifact.METHODS:Patients with suspected acute pulmonary embolism were examined by CTPA,and images were divided into 4 groups according to different thickness:0.6,1.0,3.0 and 5.0mm.Display clarity of pulmonary embolism arteries of the images of 4 groups was compared to evaluate the continuous level of pulmonary artery branches and the incidence rate of artifact.RESULTS AND CONCLUSION:The main pulmonary artery and lobe pulmonary artery of two lungs showed no significant difference among the 4 groups;segmental and sub-segmental pulmonary artery showed 0.6mm and 1.0mm slice thickness were better than 3.0mm and 5.0mm thickness groups (P0.001);for the 6th pulmonary artery,4 groups showed significant differences (P0.001),and the display rate of 0.6 mm was the best.The ability that 1.0mm slice thickness images showed sub-segmental pulmonary artery branch level and above was similar to the 0.6mm,and better than 3.0mm and 5.0mm groups;3.0mm and 5.0mm slice thickness groups showed continuous pulmonary artery all branches lower than 0.6mm and 1.0mm slice thickness groups.Partial volume artifacts of 39 cases with negative were 2 cases from 0.6mm group,5 from 1.0mm group,11 from 3.0mm group,17 from 5.0mm group,showing significant differences among 4 groups (P0.001).In CTPA examination,slice thickness at 3.0mm or less can observe sub-segmental pulmonary artery more accurately,while 0.6mm slice thickness can reduce the incidence of partial volume artifacts and improve images quality of CTPA.

关 键 词:层厚 X射线计算机 体层摄影术 肺动脉 造影 伪影 

分 类 号:R318[医药卫生—生物医学工程]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象