UF-1000i尿沉渣分析仪红细胞测定与显微镜检查结果的对比分析  被引量:7

Comparative analysis of the results of UF-1000i urinalysis analyzer and microscope in determination of red blood cells

在线阅读下载全文

作  者:刘然 刘佃香 李玉梅 

机构地区:[1]黑龙江省大庆市第二医院检验科,黑龙江大庆163461

出  处:《中国医药导报》2011年第11期166-167,共2页China Medical Herald

摘  要:目的:比较UF-1000i尿沉渣分析仪测定尿液红细胞与显微镜检查结果,对比两者符合率。方法:对624例尿标本同时采用UF-1000i尿沉渣分析仪测定红细胞和显微镜镜检两种方法检测。结果:UF-1000i检测尿中红细胞的阳性率比显微镜镜检阳性率高,假阳性率达到20.5%。结论:尿液标本采用UF-1000i检测红细胞时结果假阳性偏高,对此标本必须进行显微镜镜检复核。Objective: To compare the compliance rate of UF-1000i urinalysis analyzer with microscope in determination of urine red blood cells.Methods: A total of 624 cases of urine specimens were collected and examined by UF-1000i urinalysis analyzer and microscope.The results of determination of the two methods were compared and analyzed.Results: In determination of urine red blood cells,the positive rate of UF-1000i urinalysis analyzer was higher than that of microscope.The false positive rate was 20.5%.Conclusion: The false positive rate of UF-1000i urinalysis analyzer is a little higher,in determination of the red blood cells in the urine specimens,which must be reviewed by microscope.

关 键 词:UF-1000i尿沉渣分析仪 红细胞 显微镜镜检 

分 类 号:R446.12[医药卫生—诊断学]

 