造影指导下穿刺腋静脉植入起搏器电极导线  被引量:8

Placement of pacemaker leads using the axillary vein guided by contrast venography

在线阅读下载全文

作  者:段徐[1] 凌峰[1] 于忠[1] 许轶洲[1] 王振东 徐海鹰[1] 沈芸[1] 金晶[1] 

机构地区:[1]杭州市第一人民医院心血管内科,310000

出  处:《中华心律失常学杂志》2011年第1期34-36,共3页Chinese Journal of Cardiac Arrhythmias

摘  要:目的探讨造影指导下穿刺腋静脉植入起搏器电极导线的方法。方法2009年5月至2010年4月,共145例患者采用造影指导下穿刺腋静脉植入电极导线。经前臂静脉注射造影剂,记录腋静脉影像,以此为指导,行腋静脉穿刺。结果穿刺腋静脉植入起搏器电极导线成功率为91.6%,穿刺成功组与穿刺不成功组平均年龄、性别比例差异无统计学意义,1例(0.7%)患者出现气胸,1例(0.7%)因误穿腋动脉并插入扩张管及外鞘导致皮下广泛出血,无电极导线磨损、断裂。结论造影指导下穿刺腋静脉植入起搏器电极导线安全、有效,是起搏器电极导线植入的重要方法。Objective To evaluate the safety and efficacy of placement of pacemaker leads using the axillary vein guided by contrast veuography. Methods Pacemaker leads were placed using axillary vein guided by contrast venography in 145 patients during May 2009 to April 2010. Contrast dye was injected via a peripheral vein for recording venography image of axillary vein, and the axillary vein was punctured guided by the venog- raphy image. Results The success rate of placement of pacemaker leads through axillary vein guided by con- trast venography was 91.6%. There was no significant difference between the success group and no success group in average age and sexual ratio. One patient(0. 7% )developed insignificant pneumatothorax, one patient (0. 7% )developed significant subcutaneous hemorrhage caused by puncturing into artery by mistake. There was no patient developed lead abrasion or fracture. Conclusion Placement of pacemaker leads using the axillary vein guided by contrast venography was safe and effective. It' s an important method for placement of pacemaker leads.

关 键 词:腋静脉 造影 穿刺 起搏器电极导线 

分 类 号:R541.7[医药卫生—心血管疾病]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象