计数法测定脏器容积  被引量:1

Determination of organ volume by counting methods

在线阅读下载全文

作  者:何淑杰[1] 张云平 张萌萌 张英男[1] 孝延龄[1] 吕丽莎[2] 赵文举 

机构地区:[1]白求恩医科大学第三临床学院,长春130021 [2]长春市中心医院,长春130051

出  处:《核技术》1999年第11期678-681,共4页Nuclear Techniques

摘  要:利用三组模型研究SPECT脏器功能容积测定中衰减校正系数、放射性活度、容积大小、放射性缺损区和阈值对容积测定的影响。结果表明,在SPECT图像重建过程中,不进行衰减校正测得的容积明显高于模型实际容积,测得的容积随校正系数的增大而减小。当校正系数为0.1时,测得的容积与模型实际容积最接近。放射性活度在37—185kBq范围内时,测得的容积没有明显差别(P>0.05)。阈值增加,测得容积减小;当阈值为48%时,测得的容积与模型实际容积最接近。放射性“冷区”的存在对功能容积的测量无影响。The effects of attenuation correction coefficient, radioactivity, volumedimensions, radio deficit area and threshold on the measurement of organ volumewere investigated using 3 groups of phantom. The results showed that the measuredvolume without attenuation correction was higher than the real volume of phantomsin the reconstruction of SPECT images. The measured volume was getting smaller asthe correction coefficient increased. When correction coefficient' was 0.1, themeasured volume approximated mostly the real volume of phantom. There was nosignificant change in the measured volumed when radioactivity ranged from 37 to185kBq/mL (P>0.05). The measured volume decreased as threshold increasing andbest results were obtained when the threshold was 48%. The presence of radio deficitarea showed no influence for the determination of functional volume. Conclusions:the best results can be obtained with Change's attenuation correction method inreconstruction with correction coefficient 0.1 and threshold 48%.