机构地区:[1]河北医科大学第四医院,河北省石家庄市050011 [2]中国疾病预防控制中心
出 处:《中国全科医学》2011年第13期1428-1430,1437,共4页Chinese General Practice
摘 要:目的分析血压在短时间内重复测量中的波动规律,比较汞柱式血压计和全自动电子血压计波动规律的差异。方法利用Y型三通技术,用汞柱式血压计与电子血压计(OMRON HEM-770A)同步测量85名受试者,以汞柱血压作为参考血压分析间隔1~2 m in三次测量结果的波动规律,并与电子血压进行比较。结果三次测量所得参考血压存在规律波动,第二次比第一次平均低3.5/1.8 mm Hg;第三次比第二次低1.8/0.8 mm Hg,后两次的血压波动趋于平稳。与第一次比较,第二次测量95%的个体收缩压波动在-16~9 mm Hg之内,舒张压在-12~8 mm Hg之内;与第二次比较,第三次测量95%的个体收缩压波动在-14~11 mm Hg之内,舒张压在-8~6 mm Hg之内。电子血压计各次读数间也存在相似波动或偏移。结论人的血压可存在超过10 mm Hg的瞬时波动,电子血压计能较好的记录血压的瞬时变化,不能因两次测量读数不同而轻易怀疑血压计的稳定性。Objective To explore the fluctuation rules of blood pressures in repeated measurements during a short period of time by using mercury sphygmomanometers and automatic electronic ones,and then compare the difference.Methods The blood pressures of 85 participants were measured simultaneously with mercury sphygmomanometer(blood pressure as a reference) and electronic blood pressure monitor(OMRON HEM-770A) through a Y-joint device for three times with a 1-to 2-minute interval.Results A regular fluctuation was found in blood pressures as reference gained in 3 times of measurement: the pressure in the second measurement was lower than that in the first one by average 3.5/1.8 mm Hg;the pressure in the third measurement was 1.8/0.8 mm Hg lower than that in the second one,and the pressures in the last two measurements were closer.Compared with the first measurements,in the second measurement,the systolic blood pressures fluctuated between-16 mm Hg and 9 mm Hg,and the diastolic blood pressures changed between-12 mm Hg and 8 mm Hg in 95% individuals.Compared with the second measurements,95% of the third-time measured systolic blood pressures fluctuated between-14 mm Hg and 11 mm Hg,and diastolic pressures changed between-8 mm Hg and 6 mm Hg.The results measured with the electronic device showed the similar fluctuation or deviation.Conclusion At least 10mmHg fluctuation may exits between repeated blood pressure measurements in a short period of time.People should not doubt the accuracy of electronic sphygmomanometers just according to the difference between two repeated measurements.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.31