nano-TiO_2修饰玻碳电极微分脉冲阳极溶出伏安法测定全血中砷含量  

Determination of arsenic by using differential pulse anodic stripping voltammetry on the nano-TiO_2 film modified glass-carbon electrode

在线阅读下载全文

作  者:黄亚北[1] 许希海[1] 丁新平[1] 王士军[1] 訾言勤[2] 

机构地区:[1]淮北职业病防治院,安徽淮北235000 [2]淮北师范大学,安徽淮北235000

出  处:《中国卫生检验杂志》2011年第5期1100-1102,共3页Chinese Journal of Health Laboratory Technology

摘  要:目的:建立一种电化学测定全血中砷的方法。方法:采用微波消解法处理全血样品,在优化测定条件下以微分脉冲阳极溶出伏安法测定砷的含量。结果:砷的检出限为0.37μg/L;砷标准溶液在1.0μg/L^15.0μg/L范围内线性良好,相关系数为0.998;测定含5.0μg/L As标准溶液的相对标准偏差为2.0%。样品加标回收率为95%~97%。结论:该方法灵敏度、准确度高;操作方便快速;具有较低的检出限,能满足测定全血中砷含量的检测工作。Objective:To establish a method for simultaneous determination of arsenic in blood by electrochemical method with microwave digestion.Methods: Under the proposed conditions,arsenic was determined in blood by differential pulse anodic stripping voltammetry(DPASV) technique with microwave digestion.Results: The detection limits were 0.37 μg/L for arsenic.The liner calibration curves for arsenic was respectively with in the range of 1.0 μg/L^15.0 μg/L,and their correlative coefficients was 0.998.The RSD for arsenic at 5.0 μg /L was 2.0%.The standard recovery rates of spiked samples were in the range of 95%~97% for arsenic.Conclusion: The method is simple,rapid,accurate,sensitive and of low detection limits.DPASV technique,new,rapid,simple,selective and inexpensive for qualitative and quantitative determinations of arsenic,which was successfully applied.

关 键 词:微分脉冲阳极溶出伏安法 修饰电极 微波消解  

分 类 号:O657.13[理学—分析化学]

 