对CIBA-CORNING 644型分析仪定标和斜标液的研制  被引量:2

Develop of calibrating and stope standard solution of using the CIBA CORNING 644 model ion selective electrode analyzer

在线阅读下载全文

作  者:王广结[1,2,3] 莫德格 王钢[1,2,3] 王伟中 张军力[1,2,3] 

机构地区:[1]内蒙古血液中心 [2]呼和浩特市第一医院检验科 [3]内蒙古医学院附属医院检验科

出  处:《内蒙古医学杂志》1999年第4期197-199,共3页Inner Mongolia Medical Journal

摘  要:目的:对644型离子选择电极分析仪所用定标和斜标液进行研制。方法:采用644型离子选择电极分析法。结果:①进口与自配浓度一致,两组无显著性差异。②重复性试验:批内CV值小于10%,批间CV值小于20%。③回收率试验:Na+、K+、Cl-均值分别为101%、988%、1003%。结论:自配试剂的精密度和准确度非常满意,符合要求。objective:Develop the calibrating and stope standard solution of using the CIBA CORNING 644 model ion selective electrode Na +/K +/Cl analyze.Methods:use the 644 model ion selective electrode analyze method.Results:The concentration of calibrating and stope standard solution of imported and self developed was consistent.There was no obvious difference between two groups.The repetition tests showed that:the CV of intra group was<1.0%, the CV of inter group was <2.0%.The recovery rate tests showed that:Na +:101%、K +:98.8%、 Cl -:100.3%.Conclusion:The precision and degre of accuracy in reagent of self developed was feel very satisfied.accord with the demands. [

关 键 词:离子选择电极 定标液 斜标液 临床检验 

分 类 号:R446[医药卫生—诊断学]

 