RIA检测HBsAg后带现象的初步探讨  被引量:1

Preliminary Stuty of Postzone Phenomenon in HBsAg detecton Used by RIA

在线阅读下载全文

作  者:沈继录[1] 刘霞秋[1] 李正华[1] 

机构地区:[1]安徽省蚌埠医学院附属医院核医学科,蚌埠233004

出  处:《标记免疫分析与临床》1999年第4期218-220,共3页Labeled Immunoassays and Clinical Medicine

摘  要:为探讨放免法检测HBsAg 是否存在后带现象, 采用放射免疫分析(RIA) 技术, 用二步法和一步法对595 例标本进行了检测。结果显示,用二步法测定HBsAg 为强阳性的标本, 用一步法测定时有部分标本变为弱阳性和假阴性。对此部分标本进行1:10 和1 :100 稀释后, 再用一步法测定时全部变为强阳性。提示一步法检测HBsAg 时, 也存在后带现象。The article is to study the postzone phenomena in HBsAg detection used with RIA.Using RIA, 595 samples were tested with two step method and one step method. The results showed that the strong positive samples of HBsAg in detection using with two step method partly became slight positive and false negative, when these samples were detected by one step method. After these were diluted into 1:10 and 1:100, these were strong positive with one step method. The study indicated that HBsAg detection used with one step method had postzone phenomena too.

关 键 词:放射免疫分析 乙型肝炎 HBSAG 后带现象 