作 者:刘琼[1,2] 高云涛[1,2] 放茂良[1,2] 李东东[1,2] 李乔丽[1,2]
机构地区:[1]云南民族大学化学与生物技术学院,昆明650031 [2]民族药资源化学国家民委-教育部共建重点实验室(云南民族大学),昆明650031
出 处:《药物分析杂志》2011年第7期1402-1406,共5页Chinese Journal of Pharmaceutical Analysis
摘 要:目的:建立嵌入式碳纳米管玻碳电极测定微量铅测定中的应用。方法:把碳纳米管膜嵌入到玻碳电极表面,用其作为工作电极并与铋膜法相结合,采用微分电位溶出法测定微量铅的含量。结果:与玻碳电极相比,碳纳米管修饰电极能明显提高灵敏度和稳定性。三七根、竹叶兰、青竹标中均含一定量的铅。方法的最低检出浓度为0.6000μg.L-1,加标回收率93.9%~108.8%,RSD为2.79%~8.60%。结论:嵌入式碳纳米管玻碳电极具有准确、高灵敏度、稳定性更高等特点,能被用于其它微量元素的测定。Objective:To establish a novel differential potentiometric stripping analysis method to detect trace lead in using inlaid CNT-modified electrode.Methods:Carbon electrode was modified by inlaid CNT-modified electrode and stripping.Results:The sensitivity was improved obviously with inlaid CNT-modified electrode compared with bared carbon electrode,and stability improved obviously.They contains some lead element.The experimental results showed that the detection limits was 0.6000 μg·L-1,the addition standard recovery(ASR)was 93.9%-108.8%,and the RSD was 2.79%-8.60% for Pb.Conclusion:Inlaid CNT-modified electrode is accurate,high sensitivity,stability higher etc,can be used in the determination of trace element.
关 键 词:嵌入式 碳纳米管 铋膜 微分电位溶出 铅 民族药 三七根 竹叶兰 青竹标
分 类 号:R917[医药卫生—药物分析学]