摘 要:目的:探讨血细胞分析仪测定血小板结果的准确性,分析错误产生的原因。方法:用Sysmex XE-2100血细胞分析仪测病人的血小板,选择血小板直方图异常的病例二组,第一组血小板计数低于80×10~9/L的病例80例、第二组血小板计数高于500×10~9/L的病例40例,用显微镜计数法复查血小板结果。结果:在第一组病例中,两法测得血小板结果有显著性差异(P〈0.001);在第二组病例中,两法测得血小板结果也有显著性差异(P〈0.05)。结论:实际工作中不能完全依赖血细胞分析仪测得的结果,对于异常结果还须用传统的显微镜法进行复查。Objective:To explore the accuracy of platelet results tested with blood cell analyzer and analyze the causes of error produced.Methods: Test platelet of patient with Sysmex XE-2100 blood analyzer.Choose two groups of cases which platelet flags were abnormal among them. The first group included 80 cases(PLT80×109/L),the second group included 40 cases(PLT500×109/L).Review platelet results with Microscope counting method.Results:The results tested by two methods quite disagreed in the first group(P0.001).There was significant difference in the second group too(P0.05).Conclusion:In practical work medical laboratory technicians can't rely completely on measurement results with blood cell analyzer and still must review abnormal results with the traditional microscope counting method.