ADVIA-2120血液分析仪测定有核红细胞的评估  

Evaluation of ADVIA-2120 Hematology Analyzer for Measuring Nucleated Red Blood Cell

在线阅读下载全文

作  者:梁小亮[1] 

机构地区:[1]厦门市第二医院检验科,福建厦门361021

出  处:《医学临床研究》2011年第7期1310-1312,共3页Journal of Clinical Research

摘  要:【目的】评价SIEMENSADVIA-2120型血液学分析仪对有核红细胞(NRBC)分析结果的准确性。【方法】采用仪器法和手工法分别对120份外周血中含有有核红细胞的临床血液样本进行测定。【结果】仪器法和手工法对NRBC的检测结果比较,中、高值差异无统计学意义(P〉0.05),低值差异有统计学意义(P〈0.01)。标本重复性能良好,平均变异系数(CV)〈6%,灵敏度92.3%,特异性97.7%,假阳性14.3%,假阴性1.16%。【结论】ADVIA-2120血液分析仪测定NRBC是一种较为经济、简洁、准确的方法,适宜在临床疆产使鼠,[Objective] To evaluate the accuracy of the results of nucleated red blood celI(NRBC) measured by SIEMENS ADVIA-2010 hematology analyzer. [Methods]The peripheral blood NRBC of 120 patient s were measured by both instrumental method and manual method. [ResultslThere was no significant difference in the high value and medium value of the measurement results between the instrumental method and manual method( P 〉0.05), but there was significant difference in the low value between two methods( P 〈0.01). The repetitiveness of the sample was good. The average CV was below 6%. Sensitivity and specificity was 92.3% and 97.7%, respectively. The false positive rate and false negative rate was 14.3 % and 1.16 %, respectively. [ConelusionlADVIA-2120 hematology analyzer for measuring NRBC is more economic, convenient and accurate method. Therefore, it is suitable to be extended in the clinics.

关 键 词:血液化学分析/仪器和设备 红细胞 

分 类 号:R446.112[医药卫生—诊断学]

 