一等标准水银温度计检定结果持续有效性保证方法研究  被引量:1

Research on Procedure to Assure the Continued Validity of the Verification Results of Standard Mercury-in-Glass Thermometer(Grade I)

在线阅读下载全文

作  者:周娟[1] 刘光荣[1] 严勇[1] 曹德森[1] 吴昊[1] 

机构地区:[1]解放军总医院医学保障部医学工程保障中心,北京100853

出  处:《生物医学工程学杂志》2011年第4期721-724,731,共5页Journal of Biomedical Engineering

摘  要:一等标准水银温度计可用来完成温度的计量溯源。为保证检定结果的持续有效性,一等标准水银温度计的完全重校准是必要的。文章对一等标准水银温度计零位值的完全重校准方法进行了研究,指出目前国内恒温槽实际温度计算方法与国外的不同,并对两种计算方法中涉及的分度修正值、可调分度修正值进行了比较。Standard mercury-in-glass thermometer(Grade I) can be used to perform traceable measurements.Full recalibration is necessary to assure the continued validity of the verification results of the standard mercury-in-glass thermometer(Grade I).The present paper shows researches on procedure for the recalibration at the ice point of standard mercury-in-glass thermometer(Grade I),and points out different calculation ways of the true temperature of the thermostatic bath.The different values of scale correction and adjusted scale correction are compared in this paper.

关 键 词:检定结果的持续有效性 完全重校准 恒温槽实际温度 分度修正值 可调分度修正值 

分 类 号:TH811[机械工程—仪器科学与技术]

 