床旁监测血糖仪葡萄糖氧化酶法及脱氢酶法准确度评估  被引量:1

Accuracy of blood-glucose monitoring system by using glucose oxidase or glucose dehydrogenase

在线阅读下载全文

作  者:邸玉玮[1] 郑卫东[1] 梁雪冰[1] 戴耀宗[1] 

机构地区:[1]广东省医学科学院,广东省人民医院检验科,广州510018

出  处:《中华内分泌代谢杂志》2011年第8期634-635,共2页Chinese Journal of Endocrinology and Metabolism

摘  要:以生化分析仪检测为参照,比较葡萄糖氧化酶法及脱氢酶法血糖仪的准确度。检测新鲜肝素锂化静脉血临床样本。根据ISO15197,血糖仪的两种方法检测结果偏倚100%在±0.56mmol/L内(〈4.2mmol/L),≥99.7%在±20%内(≥4.2mmol/L);根据EF9-A2,两种方法的预期误差〈10%,略〉5%;低、高血糖症界值处(〈3.89mmol/L或〉6.1lmmol/L),两种方法的κ〉0.6。两种方法间的κ〈0.6。血糖仪葡萄糖氧化酶法、脱氢酶法准确度符合ISO15197要求,与生化分析仪检测方法的差异不影响临床应用,其结果用于低、高血糖症判读一致性好。Blood samples were detected by 24 blood-glucose meters using glucose oxidase and another 18 glucose meters using glucose dehydrogenase. The plasma glucose was detected by the auto-chemistry analyzer as control. According to ISO15197 and EF9-A2, the bias of results from both glucose meters 100% fall in the range of ± 0. 56 mmol/L( 〈4. 2 mmol/L)and ≥ 99. 7% in the range of ±20% ( ≥4. 2 mmol/L) , predicted bias were all less than the true bias(Bc). As κ〉0. 6, the results from both glucose meters were in accordance with the results from auto- chemistry analyzer in judging hyperglycemia( 〉6. 11 mmoL/L) and hypoglycemia( 〈3.89 mmol/L). Between the results from the two blood-glucose meters, κ〈0. 6. The accuracy of both glucose-meters are accepted for the purpose of clinical diagnosis and treatment.

关 键 词:血糖 测定 准确性 

分 类 号:R446.1[医药卫生—诊断学]

 