作 者:吕俊霞[1] 祁晓婷[2] 徐卫益[1] 陈保德[1] 毛卫林[1]
机构地区:[1]浙江大学医学院附属第一医院检验科,浙江杭州310003 [2]新疆维吾尔自治区人民医院特检科
出 处:《实验与检验医学》2011年第4期363-365,共3页Experimental and Laboratory Medicine
摘 要:目的分析、比较三台血液分析仪阻抗法与显微镜手工法计数低值血小板结果,并探讨其对血小板输血阈值判断的影响。方法选择本实验室三台血液分析仪(LH750、CD-3700和Sysmex-xt1800i)分别计数低值血小板患者274例,PLT均小于100×109/L,以显微镜手工法为参考,按检测结果将血小板分为0<PLT≤5×109/L,5×109/L<PLT≤10×109/L,10×109/L<PLT≤20×109/L,20×109/L<PLT<100×109/L四组,并与手工法结果进行比较分析。结果三台血液分析仪与镜检法计数在PLT≤20×109/L时相关系数(γ)均<0.8。当20<PLT<100×109/L时相关系数(γ)均≥0.8.从ROC曲线分析,三台血液分析仪分别在0<PLT≤5×109/L和5×109/L<PLT≤10×109/L两组中AUC均小于0.90,在10×109/L<PLT≤20×109/L组中AUC均大于0.90(分别为0.913,0.904,0.947)。结论三台血液分析仪计数PLT≤20×109/L的样本与镜检法相关性较低或无直接相关性,故对低值血小板,考虑需要血小板输血时,应慎重对待仪器检测的结果。由ROC曲线可知Sysmex-xt1800i对不同输血阈值的判断要优于LH750和CD-3700。Objective To compare low PLT counts produced by impedance(automated blood analyzer) and microscopic methods and explore their effect on determining to platelet transfusion.Methods A total of 274 samples with low PLT count(100×109/L) were tested by three automated blood analyzers(LH750,CD-3700 and Sysmex-xt1800i),and then were divided into four groups(0PLT≤5×109/L,5×109/LPLT≤10×109/L,10×109/LPLT≤20×109/L and 20×109/LPLT100×109/L) according to the microscopic method.The results of these two methods were compared.Results The correlation coefficient(γ) of two methods were less than 0.8 when platelet counts were 0×109/L-20×109/L,if platelets were 20×109/L-100×109/L,the correlation coefficient(γ) were more than 0.8.from the ROC cure analysis,three antomated blood analyzers′s ROC curve of areas were less than 0.90 when platelets were between 0×109/L-5×109/L and 5×109/L-10×109/L.Areas were more than 0.90 if platelets were 10×109/L-20×109/L(respectively 0.913,0.904,0.947).Conclusions Three automated blood analyzers are not directly related to microscopic method when platelets were 0×109/L-20×109/L.For the low platelets,considering the need for platelet transfusions,the patients should be treated with the instrument test′s results strictly.ROC curve indicated the Sysmex-xt1800i was the best in the three automated blood analyzers for the judgments of transfusion thresholds.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.211