机构地区:[1]首都医科大学附属北京友谊医院放射科,北京100050
出 处:《中国医学影像技术》2011年第10期2146-2148,共3页Chinese Journal of Medical Imaging Technology
摘 要:目的应用CDRAD 2.0模体对数字影像(DR)的密度分辨力进行定量评价。方法用两台Philips DR机对CDRAD 2.0模体进行数字化成像,设备1投入使用时间为7年,设备2使用时间为17个月。用匹配软件处理数据资料,计算IQFinv值,比较两台设备的密度分辨力差异。结果设备1的IQFinv均值为2.31,设备2的IQFinv均值为2.81,二者差异有统计学意义(t=7.855,P<0.01)。结论采用CDRAD 2.0模体可定量分析DR设备的密度分辨力。Objective To evaluate the density resolution of digital radiology(DR) system subjectively by using CDRAD 2.0 phantom.Methods Digital radiograms of CDRAD 2.0 phantom were obtained by two Philips DR system(Unit 1 and Unit 2),the using time of which was 7 years and 17 months,respectively.The image data were analyzed with the processing software of the phantom.IQFinv value defined in a formula was calculated of each modality.Results IQFinv value of Unit 1 was 2.31,that of Unit 2 was 2.81.There was statistical difference between the two units(t=7.855,P0.01).Conclusion The density resolution of DR can be quantitatively evaluated using CDRAD 2.0 phantom.