LH750自动血细胞分析仪的性能初步评价  被引量:8

Performance Evaluation of LH 750 Automatic Hematology Analyzer

在线阅读下载全文

作  者:乐家新[1] 兰亚婷[1] 徐菡[1] 炼鑫[1] 王红霞[1] 朱红[1] 

机构地区:[1]解放军总医院临床检验科

出  处:《医疗卫生装备》2011年第10期86-88,共3页Chinese Medical Equipment Journal

摘  要:目的:评价LH 750自动血细胞分析仪全血细胞计数(CBC)和白细胞分类(DC)的性能。方法:按照有关文件规定对该仪器CBC的性能指标进行评价,并与人工分类进行比较。结果:该仪器CBC的精密度、携带污染率和总重复性均符合要求;用LIN-C线性质控物进行测定,结果均在期待值范围;与对照仪器的结果比较,相关性好(r^2〉0.963 0),偏倚%小(-2.094-1.284)。DC:中性粒细胞(Neut)、淋巴细胞(Lym)、单核细胞(Mon)、嗜酸性粒细胞(Eos)和嗜碱性粒细胞(Baso)的重复性好;WBC形态正常的样本,仪器与人工DC的结果比较,Neut、Lym和Eos相关性较好(r^2=0.946 9-0.966 1);Mon次之(r^2=0.801 3),Baso的相关性欠佳(r^2=0.518 0)。白细胞分类报警的假阳性为15.93%,假阴性为4.35%,准确度为86.78%,特异度为84.07%,敏感度为95.65%。结论:该仪器检测的结果准确、可靠,主要性能指标符合实验要求;对怀疑WBC分类异常的标本,仪器DC结果必须经显微镜复核后才能报告。Objective To evaluate the performance of LH 750 hematology analyzer.Methods The indexes including precision,total reproducibility,carry-over,linearity and comparability were examined according to published evaluation standard.At the same time,the indexes of differential count(DC) were compared between those produced by the analyzer and manual methods.Results Precision,carry-over and total reproducibility in counting of leucocytes,erythrocytes,platelets and examination of hemoglobin on the analyzer were showed in the prescriptive range.The determined results for linearity-control were expected ranges.The results correlated well(r^2〉0.963) and bias were little(-2.094%-1.284%) with those obtained on the comparative instrument.DC:reproducibility was good for neutrophils,lymphocytes,monocytes,eosinophils and basophils.Comparison of the results by instrument with manual for normal samples in morphology,the correlation was better for neutrophils,lymphocytes and eosinophils(r^2=0.946 9-0.966 1);good for monocytes(r^2=0.801 3),not good for basophils(r^2=0.518 0).Comparison of instrument flags with manual DC for 296 samples,false positive ratio,false negative ratio,accuracy,specificity and sensitivity were 15.93%,4.35%,86.78%,84.07% and 95.65% separately.Conclusion The analyzer can be used to test blood cell parameters accurately and reliably.Its main performance indices accords with the experimental requirements;the results are credible.It is necessary to check with microscopy for DC before reported when it is doubted.

关 键 词:血细胞分析仪 精密度 重复性 携带污染率 线性 

分 类 号:R446.113[医药卫生—诊断学] R318.6[医药卫生—临床医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象