石墨炉原子吸收光谱法测定地下水中铅不确定度传递模型研究  被引量:1

Research on Uncertainty Propagation Model of Determination of Lead in Groundwater by Graphite Furnace Atomic Absorption Spectrometry

在线阅读下载全文

作  者:郑荣华[1] 刘建坤[1] 朱家平[1] 

机构地区:[1]南京地质矿产研究所,南京210016

出  处:《现代科学仪器》2011年第5期118-120,共3页Modern Scientific Instruments

基  金:国土资源部公益性行业科研专项资助(NO.200811133)

摘  要:利用标准曲线各点标准值X、响应值Y及其不确定度进行双误差回归,解决了标准曲线回归对不确定度的影响。结合其它影响因素,对石墨炉原子吸收光谱法测定地下水中铅的测量不确定度进行评定,建立本方法的不确定度传递模型。采用本模型,可得出未知水样包含不确定度的测定结果,给定误差下的方法检出限并与传统检出限进行了比较。Using standard points X,responses Y and their uncertainties conducted double error regression,the effect of standard curve regression for uncertainty was resolved.Combined with other influence factors,uncertainty of the determination of lead in groundwater by graphite furnace atomic absorption spectrometry was evaluated and the uncertainty propagation model was built.Adopting this model can get measurement result with uncertainty,methodological detection limit that given error compared with conventional detection limit.

关 键 词:石墨炉原子吸收光谱法  双误差 不确定度传递模型 

分 类 号:O657.31[理学—分析化学] X832[理学—化学]

 