作 者:姜海滨[1] 黄新苗[1] 白元[1] 陈翔[1] 宗书峰[1] 赵仙先[1] 秦永文[1] 刘宗军[2]
机构地区:[1]第二军医大学附属长海医院心内科,上海200433 [2]上海中医药大学附属普陀区中心医院心内科
出 处:《中国介入心脏病学杂志》2011年第5期268-272,共5页Chinese Journal of Interventional Cardiology
基 金:国家高科技研究发展计划(863计划;2006AA02Z4D7)
摘 要:目的测量在不同起搏频率下主动脉跨瓣压力阶差的变化,为经导管主动脉瓣置入手术确定适合的起搏参数。方法选取6只成年健康山羊,全麻下经股静脉将临时起搏电极送至右心室,连接临时起搏器,频率范围由150次/分递增至350次/分,以25次/分递增频率,测量在不同起搏频率下主动脉瓣上和瓣下的收缩压、舒张压和平均压,并计算压力阶差。结果收缩期主动脉跨瓣压力阶差随起搏频率的加快而升高,至250~300次/分时,由(-25.00±3.90)mm Hg上升至(11.17±10.85)到(13.50±3.89)mm Hg之间。舒张期跨瓣压力阶差随起搏频率的加快而降低,至250~300次/分时,稳定在(47.00±3.03)mm Hg到(43.83±2.93)mm Hg之间,压力阶差降幅达50.11%。结论 250~300次/分的起搏频率可有效地降低主动脉跨瓣压力阶差,既能保证动物的安全又可减少因压力阶差引起的瓣膜置入过程中的移位。Objective To study the changes of aotic transvalvular pressure gradient to determine the appropriate pacing frequency for transcatheter aortic valve implantation.Methods Six healthy adult goats were randomly chosen for the study were tested under general anesthesia.A temporary pacing lead connected with a temporary pacemaker was delivered to the right ventricle through the femoral vein.The pacing frequency range from 150 beats/min to 350 beats/min by increment of 25 beats/min.The aortic systolic,diastolic and mean pressure were measured at different pacing frequency and the pressure gradients in aortic were calculated.Results The transaortic valve systolic pressure gradient rose with the acceleration of the pacing rate.When the pacing rate accelerated to 250~300 beats/min,the pressure gradient rose from(-25.00±3.90) mm Hg to(11.17±10.85)-(13.50±3.89) mm Hg.Diastolic transvalvular systolic pressure gradient declined with the acceleration of the pacing rate.When the pacing rate accelerated to 250-300 beats/min,the pressure gradient was in the(43.83±2.93)-(47.00±3.03) mm Hg,showing a lower decline rate of 50.11%(P0.01,95% CI).Conclusions The pacing rate at the range of 250 beats / min to 300 beats / min is appropriate to reduce transaortic valve pressure gradients during the transcatheter aortic valve implant procedure in goats.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.3